Equipment and Calibration History

View Equipment |

View history entry 6547 for YSI EXO fDOM sn: 22E100649

Date: 2024-01-23
From: Lab - none
To: Lab - none
Title: Calibration
Notes: fDOM (QSU and RFU), Last cal: 2023-08-17 EXO sn 22C103416 EXO/Sensor Firmware: 1.0.86/3.0.0 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,-2.68 QSU,0.00 QSU,0.00 QSU,20.692 °C Pt2,300.00 QSU,182.33 QSU,300.08 QSU,0.00 QSU,21,776 °C Notes: Fresh calibration solution after a factory reset. Previous calibration attempts in the last few days were all really off. Keep an eye on this sensor. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,0.24 RFU,0.00 RFU,0.00 RFU,20.081 °C Pt2,100.0 RFU,RFU,122.16 RFU,100.05 RFU,0.00 RFU,22.019 °C Notes: Fresh calibration solution after a factory reset
Entered by: Daphne Szutu