Equipment and Calibration History

View Equipment |

View history entry 6497 for Licor LI-7500A sn: 75H-2180

Date: 2024-01-03
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2023-09-18 AGC=101.1 SDM=0 DAC1=- DAC2=- AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.8kPa, Tair=23.1C, Tdew=7.9C Zero Air at 500psi, AMP CA04119 524.77 at 700 Tdewspan=20.0 CO2 zero 0.9143 -> 0.9141 Offset: -0.8ppm CO2 span 1.0384 -> 1.0406 Offset: -1.17ppm H2O zero 0.8944 -> 0.8965 Offset: 12.9ppm H2O span 1.0344 -> 1.0383 Offset: -0.08ppm Notes:
Entered by: