View history entry 6481 for Licor LI-7810 sn: TG10-01751

Date: 2023-12-01
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2023-08-22 Zero Air at ???psi, AMP CA04119 4223.8 ppb CH4 at ???psi CO2 zero Offset: -0.024ppm H2O zero Offset: -10.52ppm CH4 zero Offset: 0.129ppb CH4 span Offset: -1ppb Notes: Calibration in the lab before Kuno's field campaign next week. Ran zero and span gases at our usual flow rate of 2 SCFH and plumbed the system with the recommended 30cm vent before the air inlet to vent excess air pressure and to prevent diffusion of ambient air into the gas stream. I waited until the slope of H2O was <0.002ppm to set the zero. I waited until the slope of CH4 was <0.002ppb to set the span. Plan 45min+ for H2O zero to stabilize. I forgot to write down the exact CH4 span offset, but I saved a screenshot and estimated the offset based on the graph.
Entered by: Daphne Szutu