Date: |
2023-11-03 |
From: |
Lab - |
To: |
Lab - |
Title: |
Calibration |
Notes: |
Last calibration: 2023-07-20
AGC=101.1 SDM= DAC1= DAC2=
AIU sn AIU-1039, Presure A0=58.372, A1=15.4
Ambient P=100.8kPa, Tair=25.1C, Tdew=9.77C
Zero Air at 700psi, AMP CA04119 524.77 at 800 Tdewspan=20.00
CO2 zero 0.868 -> .8678 Offset: -.77ppm
CO2 span 1.0095 -> 1.0116 Offset: -1.47ppm
H2O zero 0.8525 -> .8552 Offset: 15.8ppm
H2O span 0.9124 -> .9231 Offset: -.26ppm
Notes: Using loaner AMP 7550 sn AIU-1039. Dewpoint generator had low flow; fixed one of the condenser tubing which had a crack and flow was back to normal. Vertical orientation.
|
Entered by: |
|
|
|