Equipment and Calibration History
View Equipment |
Date: |
2023-11-03 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Maintenance |
Notes: |
The "from condenser" tubing was cracked so air flow was really low. Joe pulled out the old tubing with a pair of pliers and replaced it with fresh tubing. Air flow is better now. |
Entered by: |
Daphne Szutu |
|
|