View history entry 6395 for Licor LI-7810 sn: TG10-01751

Date: 2023-08-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: Zero Air at 900psi, AMP CA04119 4223.8 ppb CH4 at 850psi CO2 zero Offset: -20.9ppm H2O zero Offset: -93.5ppm CH4 zero Offset: 0.1ppb CH4 span Offset: 33.7ppb Notes: First calibration in the lab for sn TG10-01751. Ran zero and span gases at our usual flow rate of 2 SCFH and plumbed the system with the recommended 30cm vent before the air inlet to vent excess air pressure and to prevent diffusion of ambient air into the gas stream. It took more than 45min for H2O zero to stabilize.
Entered by: Daphne Szutu