Equipment and Calibration History

View Equipment |

View history entry 6376 for YSI EXO fDOM sn: 22E100649

Date: 2023-08-17
From: Lab - none
To: Lab - none
Title: Calibration
Notes: fDOM (QSU and RFU), Last cal: 5/8/2023 14:55:21 EXO sn 22C103416 EXO/Sensor Firmware: 1.0.86/3.0.0 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,-0.29 QSU,00.01 QSU,0.00 QSU,24.006 °C Pt2,297.97 QSU,316.12 QSU,297.95 QSU,0.00 QSU,24.617 °C ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,-0.74 RFU,0.00 RFU,0.00 RFU,24.666 °C Pt2,99.40 RFU,96.56 RFU,99.39 RFU,0.00 RFU,24.827 °C
Entered by: Daphne Szutu