Date: |
2023-08-07 |
From: |
Lab - |
To: |
Lab - |
Title: |
Calibration |
Notes: |
Calibration Location: Lab
Last calibration: 2023-01-18
RSSI=83%, Optical RH=28%
Ambient P=100.1kPa, Tair=27.2C
Zero Air at 900psi, AMP CA04119 4.2238 at 900
CH4 Zero Offset: -.003ppm
CH4 Span Offset: -.0318ppm
Notes: After calibration, flushed optics headspace with zero air until optics RH=11%. Removed used silica desiccant and put in fresh drierite.
|