Equipment and Calibration History

View Equipment |

View history entry 6331 for Licor LI-7500A sn: 75H-2667

Date: 2023-07-10
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2023-04-13 AGC=101.0 SDM= DAC1= DAC2= AIU sn AIU-1569, Presure A0=58.004, A1=15.479 Ambient P=100.4kPa, Tair=21.5C, Tdew=12.4C Zero Air at 1000psi, AMP CA04119 524.77 at 950 Tdewspan=16.00 CO2 zero 0.9417 -> .941 Offset: -2.21ppm CO2 span 1.0213 -> 1.0232 Offset: -1.37ppm H2O zero 0.8141 -> .8427 Offset: 7.9ppm H2O span 1.0317 -> 1.0367 Offset: -.09ppm Notes: Vertical orientation
Entered by: