Equipment and Calibration History

View Equipment |

View history entry 6305 for Licor LI-7700 sn: TG1-0357

Date: 2023-06-14
From: Bouldin Corn -
To: Bouldin Corn -
Title: Calibration
Notes: Calibration Location: Bouldin Corn Last calibration: 2022-11-30 RSSI=75%, Optical RH=23% Ambient P=101.1kPa, Tair=23.0C Zero Air at 1250psi, AMP CA04119 4.2238 at 1125 CH4 Zero Offset: -.012ppm CH4 Span Offset: .0238ppm Notes: Breezy. Flushed headspace to 12% optics RH after calibration and swapped in a new desiccant bottle.
Entered by:
Field notes