Equipment and Calibration History

View Equipment |

View history entry 6214 for Licor LI-7500A sn: 75H-2514

Date: 2023-03-14
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2022-03-17 AGC=101.4 SDM=0 DAC1=- DAC2=- Ambient P=99.2kPa, Tair=19.7C, Tdew=12.3C Zero Air at 1400psi, AMP CA04119 524.77 at 1250 Tdewspan=16.0 CO2 zero 0.905 -> 0.9053 Offset: 1.4ppm CO2 span 1.0125 -> 1.0169 Offset: -3.02ppm H2O zero 0.8403 -> 0.8408 Offset: 3.4ppm H2O span 1.0196 -> 1.0243 Offset: -0.9ppm Notes: Using AMP loaner AIU-2312
Entered by: