Date: |
2023-02-02 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Last calibration: 2022-03-02
AGC=101.8 SDM= DAC1= DAC2=
Ambient P=101.3kPa, Tair=17.9C, Tdew=4.5C
Zero Air at 1400psi, AMP CA04119 524.77 at 1250 Tdewspan=16.00
CO2 zero 0.9016 -> 0.9035 Offset: 7.88ppm
CO2 span 1.0075 -> 1.0257 Offset: -12.34ppm
H2O zero 0.8707 -> .8731 Offset: 16.0ppm
H2O span 1.0158 -> 1.0331 Offset: -.36ppm
Notes: First calibration in 11 months. Sensor installed on BWW urban tower for the past 10 months. Sensor removed this morning and will go back up on the BWW urban tower this afternoon. I updated the calibration coefficients directly on the BWW computer instead of uploading the new config file. I also swapped the HMP fan its screw terminal that was all corroded.
|
Entered by: |
|
|
|