Equipment and Calibration History

View Equipment |

View history entry 6134 for Licor LI-7500RS sn: 75H-0128

Date: 2022-11-15
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-09-21 AGC=102 SDM=0 DAC1=- DAC2=- Ambient P=101.1kPa, Tair=23.3C, Tdew=0.53C Zero Air at 1800psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00 CO2 zero 1.1643 -> 1.1626 Offset: -3.68ppm CO2 span 0.9978 -> 1.0087 Offset: -7.23ppm H2O zero 1.0603 -> 1.0599 Offset: -2.2ppm H2O span 1.0172 -> 1.0374 Offset: -0.44ppm Notes: Calibrated on box sn AIU-2312. This pair is kept together because they are an RS model and belong to AMP.
Entered by: