Equipment and Calibration History

View Equipment |

View history entry 6126 for Licor LI-7500A sn: 0042

Date: 2022-11-29
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2022-11-15 AGC=99.3 SDM= DAC1= DAC2= Ambient P=99.3kPa, Tair=22.8C, Tdew=2.02C Zero Air at 1750psi, AMP CA 04119 524.77 at 1325 Tdewspan=16.00 CO2 zero 0.9481 -> .9478 Offset: -.94ppm CO2 span 1.0127 -> 1.0145 Offset: -1.15ppm H2O zero 0.847 -> .8472 Offset: 1.3ppm H2O span 1.0292 -> 1.0267 Offset: .05ppm Notes: Typed in "15.99" for 16.00C span point. Seemed to be more accurate than typing in "16.00" for 16.00C span point.
Entered by: