Date: |
2022-11-28 |
From: |
Lab - |
To: |
Lab - |
Title: |
Calibration |
Notes: |
Last calibration: 2022-11-15
AGC=107 SDM= DAC1= DAC2=
Ambient P=100.5kPa, Tair=22.4C, Tdew=2.2C
Zero Air at 1750psi, AMP CA04119 524.77 at 1300 Tdewspan=16.00
CO2 zero 0.9998 -> .9968 Offset: -9.82ppm
CO2 span 1.0178 -> 1.0199 Offset: -1.37ppm
H2O zero 0.8213 -> .8221 Offset: 4.8ppm
H2O span 1.0308 -> 1.0304 Offset: 0.00ppm
Notes: SS rest from 107 to 100 after calibration. Today's calibration was with our usual box sn AIU-0674. The previous calibration was on a AMP box sn AIU-1039 and I had reset the SS from 93 to 100. Probably some difference in box coefficients that I didn't pay attention to, but the SS should be ok now.
|
Entered by: |
|
|
|