Date: |
2022-11-30 |
From: |
Bouldin Corn - Gas Analyzer CH4 |
To: |
Bouldin Corn - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: Bouldin Corn
Last calibration: 2022-06-08
RSSI=80%, Optical RH=23%
Ambient P=101.3kPa, Tair=12.6C
Zero Air at 1650psi, AMP CA04396 1.9317 at 1100
CH4 Zero Offset: .032ppm
CH4 Span Offset: .0083ppm
Notes: Breezy. After calibration we flushed headspace with zero air until optics RH = 8%. We swapped in fresh drierite. Half of the thermocouple tubing had broken off so we put on a new piece of tubing.
|