Equipment and Calibration History

View Equipment |

View history entry 6065 for Licor LI-7500A sn: 0073

Date: 2022-11-15
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2022-04-08 AGC=101.2 SDM= DAC1= DAC2= Ambient P=101.2kPa, Tair=23.2C, Tdew=1.69C Zero Air at 1850psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00 CO2 zero 0.9163 -> .9165 Offset: .68ppm CO2 span 0.9983 -> 1.0085 Offset: -7.22ppm H2O zero 0.8246 -> .8296 Offset: 28.9ppm H2O span 1.0147 -> 1.0258 Offset: -.25ppm Notes: Used AMP LI-7550 sn AIU-1039
Entered by: