Equipment and Calibration History

View Equipment |

View history entry 6064 for Licor LI-7500A sn: 0042

Date: 2022-11-15
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2022-04-19 AGC=99.1 SDM= DAC1= DAC2= Ambient P=101.2kPa, Tair=23.0C, Tdew=2.87C Zero Air at 1850psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00 CO2 zero 0.9483 -> .9481 Offset: -.73ppm CO2 span 1.0087 -> 1.0127 Offset: -2.76ppm H2O zero 0.8444 -> .847 Offset: 16.3ppm H2O span 1.0191 -> 1.0292 Offset: -.22ppm Notes: Used AMP LI-7550 sn AIU-1039
Entered by: