Equipment and Calibration History

View Equipment |

View history entry 6063 for Licor LI-7500A sn: 0041

Date: 2022-11-15
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2022-03-30 AGC=93.5 SDM= DAC1= DAC2= Ambient P=101.2kPa, Tair=22.9C, Tdew=.1C Zero Air at 1900psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00 CO2 zero 0.9969 -> .9998 Offset: 9.35ppm CO2 span 1.0189 -> 1.0178 Offset: .23ppm H2O zero 0.8214 -> .8213 Offset: -.5ppm H2O span 1.0337 -> 1.0308 Offset: .07ppm Notes: Reset SS from 93.5 to 100 before calibrating. Used AMP LI-7550 sn AIU-1039.