Date: |
2022-11-15 |
From: |
Lab - |
To: |
Lab - |
Title: |
Calibration |
Notes: |
Last calibration: 2022-03-30
AGC=93.5 SDM= DAC1= DAC2=
Ambient P=101.2kPa, Tair=22.9C, Tdew=.1C
Zero Air at 1900psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00
CO2 zero 0.9969 -> .9998 Offset: 9.35ppm
CO2 span 1.0189 -> 1.0178 Offset: .23ppm
H2O zero 0.8214 -> .8213 Offset: -.5ppm
H2O span 1.0337 -> 1.0308 Offset: .07ppm
Notes: Reset SS from 93.5 to 100 before calibrating. Used AMP LI-7550 sn AIU-1039.
|
Entered by: |
|
|
|