Equipment and Calibration History

View Equipment |

View history entry 6062 for YSI EXO fDOM sn: 21K102591

Date: 2022-11-03
From: Gilbert Tract -
To: Gilbert Tract -
Title: Calibration
Notes: fDOM (QSU and RFU), Last cal: 10/31/2022 4:14:35 PM EXO sn 22C100485 EXO/Sensor Firmware: 1.0.86/3.0.0 UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,0.51 QSU,0.01 QSU,0.00 QSU,16.967 °C Pt2,316.50 QSU,219.51 QSU,316.49 QSU,0.00 QSU,10.855 °C ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,0.12 RFU,0.00 RFU,0.00 RFU,15.878 °C Pt2,105.50 RFU,74.86 RFU,105.50 RFU,0.00 RFU,11.257 °C
Entered by: Joe Verfaillie
Field notes