Date: |
2022-11-03 |
From: |
Gilbert Tract - |
To: |
Gilbert Tract - |
Title: |
Calibration |
Notes: |
fDOM (QSU and RFU), Last cal: 10/31/2022 4:14:35 PM
EXO sn 22C100485
EXO/Sensor Firmware: 1.0.86/3.0.0
UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,1.87 QSU,0.00 QSU,0.00 QSU,16.796 °C
Pt2,316.50 QSU,218.07 QSU,316.49 QSU,0.00 QSU,10.735 °C
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,0.20 RFU,0.00 RFU,0.00 RFU,15.801 °C
Pt2,105.50 RFU,71.69 RFU,105.50 RFU,0.00 RFU,11.214 °C |
Entered by: |
Joe Verfaillie |
|
|