Equipment and Calibration History

View Equipment |

View history entry 6060 for YSI EXO fDOM sn: 22C100780

Date: 2022-11-03
From: Gilbert Tract -
To: Gilbert Tract -
Title: Calibration
Notes: fDOM (QSU and RFU), Last cal: 10/24/2022 8:08:51 AM EXO sn 22C100485 EXO/Sensor Firmware: 1.0.86/3.0.0 UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,-1.28 QSU,0.00 QSU,0.00 QSU,16.696 °C Pt2,316.50 QSU,315.87 QSU,316.54 QSU,0.00 QSU,10.444 °C ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,0.11 RFU,0.00 RFU,0.00 RFU,10.610 °C Pt2,105.50 RFU,92.32 RFU,105.51 RFU,0.00 RFU,11.275 °C
Entered by: Joe Verfaillie
Field notes