Date: |
2022-11-03 |
From: |
Gilbert Tract - |
To: |
Gilbert Tract - |
Title: |
Calibration |
Notes: |
fDOM (QSU and RFU), Last cal: 10/24/2022 8:08:51 AM
EXO sn 22C100485
EXO/Sensor Firmware: 1.0.86/3.0.0
UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,-1.28 QSU,0.00 QSU,0.00 QSU,16.696 °C
Pt2,316.50 QSU,315.87 QSU,316.54 QSU,0.00 QSU,10.444 °C
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,0.11 RFU,0.00 RFU,0.00 RFU,10.610 °C
Pt2,105.50 RFU,92.32 RFU,105.51 RFU,0.00 RFU,11.275 °C |
Entered by: |
Joe Verfaillie |
|
|