Date: |
2022-11-01 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Turbidity (FNU), Last cal: Unknown
EXO sn 22C100485
EXO/Sensor Firmware: 1.0.86/3.0.4
A factory reset was done first before this calibration was done in the lab.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 FNU,0.24 FNU,0.00 FNU,0.00 FNU,20.384 °C
Pt2,124.00 FNU,112.11 FNU,123.99 FNU,0.00 FNU,20.389 °C
Pt3,1010.00 FNU,939.77 FNU,1009.99 FNU,0.00 FNU,20.565 °C |
Entered by: |
Joe Verfaillie |
|
|