Equipment and Calibration History

View Equipment |

View history entry 6050 for Kipp & Zonen PAR-LITE sn: 030382

Date: 2022-10-31
From: Lab - none
To: Lab - none
Title: Tested
Notes: This sensor was removed because there were spikes in its data on two different mux ports. I pulled apart the splice in the cable and didn't see any problems. From 2022-10-27 to 2022-10-31 I ran an inter-comparison out the lab window with sn 010164. The relationship was tight (r2=0.9999) and the slope was about 1.06. sn 030382 had been inter-compared on the Tonzi tower recently but in the reflected position so the range was not optimal. I think both sensors are functional and should be inter-compared with a good reference in the field to determine appropriate coefficients.
Entered by: Joe Verfaillie