View history entry 5931 for Licor LI-7500 sn: 0976

Date: 2022-08-10
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2017-06-12 AGC=57 SDM=7 DAC1= DAC2= Ambient P=100.7kPa, Tair=25.8C, Tdew=15.3C Zero Air at 300psi, AMP CA04119 524.77 at 1425 Tdewspan=20.00 CO2 zero 0.8597 -> .8596 Offset: -.03ppm CO2 span 1.0077 -> 1.0102 Offset: 1.82ppm H2O zero 0.6335 -> .632 Offset: -10.5ppm H2O span 1.031 -> 1.0138 Offset: .37ppm Notes: Pulled out this old IRGA for Robert's UHI transect via ebike. The calibration was surprisingly good given we haven't used this sensor in 5 years. Calibrated using a small white eePC netbook using a Licor serial cable that plugs directly into the green terminal block inside the box. I had to set the span temperature in software to 19.91 in order for dew point value to be 20.00 after pressing "Span".