View history entry 5906 for Forerunner-Eosense eosFD sn: FD20180067

Date: 2022-07-11
From: Lab - none
To: Lab - none
Title: Tested
Notes: Plugged in and tested in the lab with another eosense cable. The chamber seems to be fine and the clock was still accurate, so I downloaded data from 2022-05-11 to 2022-06-08. I also ran a performance check which looked fine. I did not reset the reference. Average Flux: -0.2, 0.05 STDEV Average ATM: 687.6ppm, 29.6C Average Soil: 677.7ppm, 29.8C