Equipment and Calibration History

View Equipment |

View history entry 5800 for Licor LI-7500A sn: 0042

Date: 2022-04-19
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-12-09 AGC=100.4 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=25.8C, Tdew=7.5C Zero Air at 650psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.9481 -> .9483 Offset: .85ppm CO2 span 1.0091 -> 1.0087 Offset: .14ppm H2O zero 0.8452 -> .8444 Offset: -4.9ppm H2O span 1.0213 -> 1.0191 Offset: .05ppm Notes:
Entered by: