Equipment and Calibration History

View Equipment |

View history entry 5799 for Licor LI-7500A sn: 0035

Date: 2022-04-19
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2022-01-11 AGC=99.4 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=25.4C, Tdew=8.2C Zero Air at 650psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.9372 -> .9372 Offset: -.09ppm CO2 span 1.0128 -> 1.0167 Offset: -2.67ppm H2O zero 0.8986 -> .8983 Offset: -2.2ppm H2O span 1.0282 -> 1.0310 Offset: -.06ppm Notes:
Entered by: