Equipment and Calibration History
View Equipment |
Date: |
2022-04-15 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Tested |
Notes: |
Plugged it in in the lab. The power LED was blinking intermittently and the ethernet and status LEDs were off. The fuse and fuse connector were totally rotten and the inside of the sensor body had condensation and a few tsp of water. We didn't see any obvious corrosion on the other circuit boards and no obvious leaks/cracks in the sensor. We wiped up the water and will let it air dry before sending it back to Licor. |
Entered by: |
Daphne Szutu |
|
|