Date: |
2022-04-08 |
From: |
Lab - |
To: |
Lab - |
Title: |
Calibration |
Notes: |
Last calibration: 2022-01-11
AGC=100.9 SDM= DAC1= DAC2=
Ambient P=100.7kPa, Tair=29.96C, Tdew=12.0C
Zero Air at 750psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00
CO2 zero 0.8912 -> .8908 Offset: -1.86ppm
CO2 span 1.0205 -> 1.0198 Offset: .57ppm
H2O zero 0.7314 -> .7308 Offset: -4.2ppm
H2O span 1.0259 -> 1.0160 Offset: .22ppm
Notes: In the field, this sensor had a SS ~110. In the lab the SS was much more normal ~100. I'm not sure why there is a difference, as the config file should be the same.
|
Entered by: |
|
|
|