Equipment and Calibration History

View Equipment |

View history entry 5756 for Licor LI-7500A sn: 75H-2180

Date: 2022-03-30
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-11-30 AGC=99.9 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=22.2C, Tdew=10.1C Zero Air at 800psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.9136 -> .9141 Offset: 1.91ppm CO2 span 1.0295 -> 1.0373 Offset: 5.34ppm H2O zero 0.8924 -> .8911 Offset: -8.5ppm H2O span 1.0346 -> 1.0345 Offset: .01ppm Notes:
Entered by: