Date: |
2022-03-30 |
From: |
Lab - |
To: |
Lab - |
Title: |
Calibration |
Notes: |
Last calibration: 2021-11-30
AGC=102 SDM= DAC1= DAC2=
Ambient P=100.7kPa, Tair=21.9C, Tdew=11.1C
Zero Air at 800psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00
CO2 zero 0.9175 -> .9164 Offset: -4.45ppm
CO2 span 1 -> 1.0039 Offset: -2.8ppm
H2O zero 0.8174 -> .827 Offset: 58.5ppm
H2O span 1.0192 -> 1.0223 Offset: -.06ppm
Notes: Reset SS from 102 to 100 before the calibration. Huge H2O zero offset even after waiting 45+ min. Need to check this calibration again before deploying in field.
|
Entered by: |
|
|
|