Equipment and Calibration History

View Equipment |

View history entry 5704 for Licor LI-7500A sn: 75H-2150

Date: 2022-03-02
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-09-29 AGC=100 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=25.9C, Tdew=10.4C Zero Air at 975psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.9018 -> .9016 Offset: -.7ppm CO2 span 1.0034 -> 1.0075 Offset: -2.79ppm H2O zero 0.8689 -> .8707 Offset: 12.7ppm H2O span 1.0112 -> 1.0158 Offset: -.1ppm Notes: First calibration after changing the air pump diaphragm on the dewpoint generator. The H2O span looked stable after 12-15 minutes, but I waited for 30+ min to be sure.
Entered by: