View history entry 5658 for Forerunner-Eosense eosFD sn: FD20180067

Date: 2022-01-11
From: Lab - none
To: Lab - none
Title: Tested
Notes: Ran for a few days and it had the expected zero flux. Did a performance check, which looked good. I did not reset the reference. Average Flux: -0.1, 0.10 STDEV Average ATM: 586.4ppm, 27.5C Average Soil: 581.8ppm, 27.8C
Entered by: Daphne Szutu