Equipment and Calibration History

View Equipment |

View history entry 5644 for Licor LI-7500A sn: 0041

Date: 2021-12-16
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-09-21 AGC=100.8 SDM= DAC1= DAC2= Ambient P=100.9kPa, Tair=24.0C, Tdew=6.2C Zero Air at 1000psi, AMP CA04119 524.77 at 1650 Tdewspan=20.00 CO2 zero 0.9667 -> .9960 Offset: -2.19ppm CO2 span 1.0145 -> 1.0172 Offset: -1.79ppm H2O zero 0.8224 -> .8227 Offset: 1.5ppm H2O span 1.0228 -> 1.0335 Offset: -.23ppm Notes:
Entered by: