Date: |
2021-11-30 |
From: |
West Pond - |
To: |
West Pond - |
Title: |
Calibration |
Notes: |
Calibration Location: West Pond
Last calibration: 2021-06-01
RSSI=87%, Optical RH=18%
Ambient P=101.4kPa, Tair=18.6C
Zero Air at 100psi, AMP CA04396 1.9317 at 1750
CH4 Zero Offset: .050ppm
CH4 Span Offset: .0023ppm
Notes: Relatively still day. The zero air tank pressure may have been >100psi, but the regulator was wonky. Zero was stable for 2min after applying zero.
|
Entered by: |
|
|
|