View history entry 5592 for Licor LI-7500A sn: 0073

Date: 2021-10-16
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2021-08-31 AGC=100 SDM=0 DAC1=- DAC2=- Ambient P=99kPa, Tair=99C, Tdew=99C Licor at psi, Licor 750 at 99 Tdewspan=12 CO2 zero 0.957 -> 0.9182 Offset: 0ppm CO2 span 1.0075 -> 1.0 Offset: 0ppm H2O zero 0.7337 -> 0.8187 Offset: 0ppm H2O span 1.0078 -> 1.0094 Offset: 0ppm Notes: Factory calibration at Licor after replacement of chopper, detector and source. Unknown ambient conditions and offsets. This sensor was sent to Licor because its signal was noisy. Its parts are almost 20 years old so I asked to have it renovated. Licor said, "Inspected the unit and found it in operational condition. Replaced the chopper motor, the detector and the light source. Subjected the unit to environmental tests and calibration." Should be nearly as good as new.