Equipment and Calibration History

View Equipment |

View history entry 5588 for Vaisala HMP45A sn: U2120013

Date: 2021-11-23
From: Mayberry - RH & Temp
To: Lab - none
Title: Removed
Notes: Removed from MB and brought back to the lab for testing. AirT has been drifting downward since late October.It was replaced by HMP155 sn T2820180.
Entered by: Daphne Szutu