View history entry 5542 for Texas Electronics TE525MM sn: 65560-615

Date: 2021-10-21
From: Bouldin Alfalfa - Precipitation
To: Bouldin Alfalfa - Precipitation
Title: Tested
Notes: We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 54 tips and adjusted its set screws 1.5 turns CCW (1/2 turn of both screws is a 2-3% adjustment).
Entered by: Daphne Szutu