Equipment and Calibration History

View Equipment |

View history entry 5466 for Licor LI-7500A sn: 75H-2176

Date: 2021-09-21
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-06-07 AGC=100 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=29C, Tdew=11.4C Zero Air at 1550psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero 0.8673 -> .867 Offset: -1.56ppm CO2 span 1.0067 -> 1.0023 Offset: 3.07ppm H2O zero 0.8536 -> .8526 Offset: -5.5ppm H2O span 0.9148 -> .9091 Offset: .14ppm Notes:
Entered by: