Date: |
2021-09-08 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Last calibration: 2021-06-01
AGC=102 SDM= DAC1= DAC2=
Ambient P=100.3kPa, Tair=27.75C, Tdew=15.5C
Zero Air at 1875psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00
CO2 zero 0.8913 -> .8911 Offset: -1.15ppm
CO2 span 1.0201 -> 1.0211 Offset: .7ppm
H2O zero 0.7305 -> .7313 Offset: 5.7ppm
H2O span 1.0158 -> 1.0143 Offset: .03ppm
Notes: Showing Carlos how to calibrate 7500s
|
Entered by: |
|
|
|