Equipment and Calibration History

View Equipment |

View history entry 5436 for Licor LI-7500A sn: 0042

Date: 2021-08-31
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-05-25 AGC=99.8 SDM=0 DAC1=- DAC2=- Ambient P=99.7kPa, Tair=26.4C, Tdew=13.15C Zero Air at 1800psi, AMP CA04119 524.77 at 1700 Tdewspan=20.0 CO2 zero 0.9481 -> 0.9479 Offset: -0.78ppm CO2 span 1.0041 -> 1.0063 Offset: -1.62ppm H2O zero 0.8437 -> 0.8448 Offset: 7.1ppm H2O span 1.0149 -> 1.0163 Offset: -0.04ppm Notes:
Entered by: