View history entry 5350 for Licor LI-7700 sn: TG1-0357

Date: 2021-06-24
From: Bouldin Corn - Gas Analyzer CH4
To: Bouldin Corn - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: BC Last calibration: 2021-01-20 RSSI=80.2%, Optical RH=23% Ambient P=101.5kPa, Tair=34.5C Zero Air at 1800psi, AMP CA04119 4.2238 at 1800 CH4 Zero Offset: -0.015ppm CH4 Span Offset: -0.0338ppm Notes: This 7700 was just returned from Nick and Sabina who were using it on Staten. They calibrated it on 2021-01-20 but I don't have notes on that. Also I uploaded the TG1-0224 config file before the calibration and again after calibration but I don't think that has any affect on the calibration or measurements.