Equipment and Calibration History

View Equipment |

View history entry 5329 for Licor LI-7500A sn: 75H-2180

Date: 2021-06-15
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-03-08 AGC=99.6 SDM= DAC1= DAC2= Ambient P=100.9kPa, Tair=23.5C, Tdew=10.3C Zero Air at 150psi, AMP CA04119 524.77 at 1800 Tdewspan=16.00 CO2 zero 0.9138 -> .9138 Offset: .14ppm CO2 span 1.0231 -> 1.0281 Offset: -3.25ppm H2O zero 0.8923 -> .8901 Offset: -13.4ppm H2O span 1.0243 -> 1.0164 Offset: .16ppm Notes:
Entered by: