Equipment and Calibration History

View Equipment |

View history entry 5313 for Licor LI-7500A sn: 75H-2667

Date: 2021-06-07
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-03-01 AGC=102 SDM= DAC1= DAC2= Ambient P=99.5kPa, Tair=20.8C, Tdew=9.2C Zero Air at 250psi, AMP CA04119 524.77 at 1750 Tdewspan=16.00 CO2 zero 0.9423 -> .9422 Offset: -.43ppm CO2 span 1.0092 -> 1.0122 Offset: -2.23ppm H2O zero 0.8385 -> .8404 Offset: 11.8ppm H2O span 1.0147 -> 1.0198 Offset: -.1ppm Notes:
Entered by: