Equipment and Calibration History

View Equipment |

View history entry 5312 for Licor LI-7500A sn: 75H-2176

Date: 2021-06-07
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-03-01 AGC=99.9 SDM= DAC1= DAC2= Ambient P=100.1kPa, Tair=22.3C, Tdew=8.7C Zero Air at 200psi, AMP CA04119 524.77 at 1700 Tdewspan=16.00 CO2 zero 0.8671 -> .8673 Offset: .96ppm CO2 span 1 -> 1.0067 Offset: -4.75ppm H2O zero 0.8524 -> .8536 Offset: 7.0ppm H2O span 0.9092 -> .9148 Offset: -.13ppm Notes:
Entered by: