Equipment and Calibration History

View Equipment |

View history entry 5277 for Licor LI-7700 sn: TG1-0424

Date: 2021-05-26
From: Sherman Wetland - Gas Analyzer CH4
To: Sherman Wetland - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Sherman Wetland Last calibration: 2020-11-16 RSSI=80%, Optical RH=15% Ambient P=101kPa, Tair=27.3C AMP FB04161 at 100psi, AMP CA04119 4.223 at 2050 CH4 Zero Offset: -0.022ppm CH4 Span Offset: 0.017ppm Notes: Breezy day. Zero gas was an AMP cal gas in a baby tank that we borrowed for the day. Not much zero air in the tank but it was enough.
Entered by:
Field notes