Date: |
2021-05-26 |
From: |
Mayberry - |
To: |
Mayberry - |
Title: |
Calibration |
Notes: |
Calibration Location: Mayberry
Last calibration: 2020-11-16
RSSI=77%, Optical RH=25%
Ambient P=101.0kPa, Tair=27.8C
AMP FB03864 at 1200psi, AMP CA04119 4.223 at 2000
CH4 Zero Offset: -0.028ppm
CH4 Span Offset: -0.003ppm
Notes: Breezy day. Zero gas was an AMP cal gas in a baby tank that we borrowed for the day. After the calibration, we used the zero air to flush the headspace until optics RH = 10%. The drierite was purple, so we swapped in a new bottle with fresh drierite.
|
Entered by: |
|
|
|