Equipment and Calibration History

View Equipment |

View history entry 5268 for Licor LI-7500A sn: 0041

Date: 2021-05-25
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2020-12-02 AGC=100 SDM=0 DAC1=- DAC2=- Ambient P=100.7kPa, Tair=25.9C, Tdew=10.2C Zero Air at 350psi, AMP CA03923 474.19 at 500 Tdewspan=20.0 CO2 zero 0.9971 -> 0.9966 Offset: -1.8ppm CO2 span 1.019 -> 1.0137 Offset: 3.31ppm H2O zero 0.8209 -> 0.8206 Offset: -2.0ppm H2O span 1.0273 -> 1.0266 Offset: 0.02ppm Notes: I initially forgot to load the previous coefficients file. At that time the signal strength was reading 106.3 and I reset it to 100. Then I remembered to load the previous cal file but only the CO2 and H2O readings changed. It was calibrated with the correct coefficients.
Entered by: