Equipment and Calibration History

View Equipment |

View history entry 5204 for Licor LI-7500A sn: 0073

Date: 2021-03-15
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2020-12-08 AGC=99 SDM= DAC1= DAC2= Ambient P=100.5kPa, Tair=23.7C, Tdew=1.2C Zero Air at 400psi, AMP CA3923 474.19 at 550 Tdewspan=20.00 CO2 zero 0.9564 -> .957 Offset: 1.71ppm CO2 span 1.0079 -> 1.0077 Offset: .14ppm H2O zero 0.7323 -> .7324 Offset: -1.7ppm H2O span 1.0037 -> 1.0024 Offset: .04ppm Notes: Calibrated on software v8.8.36 with 7550 firmware v8.8.32
Entered by: